background image

Turbimax W CUS65

Installation

Hauser

7

3

Installation

3.1

Incoming acceptance, transport, storage

• Make sure the packaging is undamaged!

Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter 
has been settled.

• Make sure the contents are undamaged!

Inform the supplier about damage to the delivery contents. Keep the damaged products until the 
matter has been settled.

• Check that the scope of delivery is complete and agrees with your order and the shipping 

documents.

• The packaging material used to store or to transport the product must provide shock protection 

and humidity protection. The original packaging offers the best protection. Also, keep to the 
approved ambient conditions (see "Technical data").

• If you have any questions, please contact your supplier or your sales center responsible.

3.2

Installation conditions

3.2.1

Dimensions

a0008995

Fig. 1: 

Dimensions dependend on the sensor version (see ordering information)

D

E

B

NPT ¾

A

G1

C

Ø40

(1.57)

145

(5.71)

142

(5.59)

60

(2.36)

154

(6.06)

147

(5.79)

mm (inch)

154

(6.06)

Summary of Contents for Turbimax W CUS65

Page 1: ...BA370C 07 en 01 08 71067680 Operating Instructions Turbimax W CUS65 Turbidity sensor...

Page 2: ...Endress Hauser...

Page 3: ...3 2 Installation conditions 7 3 3 Installation instructions 10 3 4 Post installation check 11 4 Wiring 11 4 1 Connection to the transmitter 11 4 2 Post connection check 11 5 Commissioning 12 5 1 Funct...

Page 4: ...ion must only be carried out by a certified electrician Technical personnel must have read and understood these Operating Instructions and must adhere to them Before commissioning the entire measuring...

Page 5: ...ion documents No repair without completed declaration 1 5 Notes on safety icons and symbols 1 6 Notes on document symbols Warning This symbol alerts you to hazards They can cause serious damage to the...

Page 6: ...ange application A 0 to 12 g l activated sludge centrate B 0 to 40 g l return sludge C 0 to 50 g l primary sludge digested sludge D 1 to 1000 FNU wastewater effluent service water condensate boiler fe...

Page 7: ...e scope of delivery is complete and agrees with your order and the shipping documents The packaging material used to store or to transport the product must provide shock protection and humidity protec...

Page 8: ...9132 Fig 3 Orientation and installation positions with retractable assembly CUA451 The pipeline diameter must be at least 100 mm 4 if reflective materials e g stainless steel are used Install the sens...

Page 9: ...low turbidity ranges 100 FNU Make sure the sensor is completey immersed at changing water levels a0008990 Fig 4 Installation with retractable assembly Arrow flow direction of the medium Installation a...

Page 10: ...evices The sensor must be installed in an immersion assembly for immersion operation Do not use the sensor suspended freely from the cable Screw the sensor into the assembly in such a way that the cab...

Page 11: ...assembly Do not use the sensor suspended freely from the cable Prevent moisture penetrating the assembly Retractable assembly Sensor installed in flow or retractable assembly Ensure the sensor is ali...

Page 12: ...E is designated for thickened sludge To achieve representative measuring results you must use a sludge concentration of at least 1 for the first calibration point To obtain reliable calibration result...

Page 13: ...g SS 1 4571 AISI 316Ti Ordering acc to product structure Technical Information TI092C 07 en 7 2 Wall mounting Wall attachment for immersion pipes Material stainless steel 1 4301 AISI 304 order no 5150...

Page 14: ...d SXK coupling Ingress protection IP 67 Order No 51503633 Plugs SXP plug 7 pole Order No 51504027 SXK coupling 7 pole Order No 51504025 Junction box Junction box for extension of cable connection betw...

Page 15: ...Please enclose the completed Declaration of Hazardous Material and De Contamination copy the second last page of these Operating Instructions with the packaging and the transportation documents No re...

Page 16: ...by four beam pulsed light method Factory calibration 9 3 Environment Ambient temperature 20 to 60 C 0 to 140 F Storage temperature 20 to 60 C 0 to 140 F Humidity 5 to 95 Ingress protection IP 68 Versi...

Page 17: ...w required Make sure that there is a sufficient turbulence for solids with a tendency to sedimentation 9 5 Mechanical construction Design dimensions see Installation conditions Weight approx 1 kg 2 2...

Page 18: ...stallation 4 7 Measuring system 10 Installation instructions 10 M Maintenance 12 Materials 17 Maximum measured error 16 Measured value 16 Measured variable 16 Measuring range 16 Measuring system 10 O...

Page 19: ...__________________________________________ ______________________________________________________________________________________________________________ We hereby certify that this declaration is fil...

Page 20: ...www endress com worldwide BA370C 07 en 01 08 Printed in Germany FM SGML 6 0 DT 71067680...

Reviews: