TOC Analyzer CA79
Commissioning
Hauser
27
8.3.4
System
A function check on the following components can be performed in this menu:
• Starting of the pump (normal
<
or fast
<<<
)
• Switching of the valves
• Switching on of the UV reactor
• UV sensor (detection of the intensity of the UV lamp)
• Check of the sensor signals (C1/C2)
• Check of the static pressure (option)
• Leak sensor
• Resetting of the operating hours (pump)
• Resetting of the operating hours (UV lamp)
• Check of the analog outputs
• Check of the digital outputs
A0046959
18
System menu
To change the status of the pump, the valves (V1, V2, V3) and the UV lamp, the user must
press the corresponding symbol.
A chart with the following values is displayed when you press the
Graph
• Measured values C1 and C2
• Difference between C1 and C2
• Temperatures T1 and T2
Access to the service area is via the
Service
key. This area is password-protected (service
password only for service technicians).
Change to the main menu
1. Press the
Save & Exit
key.
A pop-up window appears.
2. Press the corresponding key in the pop-up window.
3. If the pump hose or the UV reactor has been changed:
Press the
Yes
key.
The operating hours are now reset.
1)
If installed and enabled in the Settings menu, an empty field is displayed otherwise.