Commissioning
TOC Analyzer CA79
24
Hauser
3. Close the pop-up window by pressing the
OK
key.
A pop-up window is displayed with the calibration results.
A0046949
13
Calibration factor pop-up window
4. Press the
Yes
key to confirm the result.
If the user chooses not to use the new calibration factor, the old calibration factor
continues to be used.
The calibration factor should be in the 0.11 - 0.21 ppb/nS range. Any deviations
should be below 2%.
Performing the calibration on a device with the "System suitability test, automated"
order option
1. Connect a calibration solution with the required concentration of sucrose to INPUT 2
(black connection).
2. Connect the water solution to INPUT 3 (blue connection).
The calibration is performed automatically with all the configured parameters
from the
Setting
→ 28 menu.
The system stops once all the measurements have been repeated.
A pop-up window is displayed with the calibration results.
A0046949
14
Calibration factor pop-up window
3. Press the
Yes
key to confirm the result.
If the user chooses not to use the new calibration factor, the old calibration factor
continues to be used.