Technical data
Thermophant T TTR31, TTR35
36
Hauser
11.5.8 Electrical safety
• Protection class III
• Overvoltage category II
• Pollution level 2
11.6 Process
11.6.1 Process temperature range
–50 to +150 °C (–58 to +302 °F)( or –50 to +200 °C (–58 to 392 °F) with extension neck).
Restrictions depending on process connection and ambient temperature:
• No restriction with compression fitting (see Accessories, order no. 51004751, 51004753)
and extension neck length min. 20 mm (0.79 in)
• with process connection:
Max. ambient temperature
Max. process temperature
up to 25 °C (77 °F)
No restrictions
up to 40 °C (104 °F)
135 °C (275 °F)
up to 60 °C (140 °F)
120 °C (248 °F)
up to 85 °C (185 °F)
100 °C (212 °F)
11.6.2 Process pressure range
Maximum permitted process pressure depending on the insertion length