SS2100 TDLAS Gas Analyzer
22
Hauser
Replacing a Fuse
1.
Power off the system and close the sample supply valve.
2.
Open the electronics enclosure. For fuse location → 5–1 (AC) or →
5–2 (DC).
3.
Using a flat-head screwdriver, remove the fuse screw turning counter
-
clockwise as shown below.
4.
Remove the fuse cover and fuse.
5.
Remove the fuse from the cover and replace with a new fuse as shown
below. Refer to Table 5–1, “Fuse specifications”.
6.
Insert the new fuse into the screw cover and replace into the fuse
opening.
Figure 5–3 Unscrewing fuse cover
FUSE COVER
Figure 5–4 Replacing fuse
FUSE