RATA (Relative Accuracy Test Audit)
The
RATA
parameter enables or disables user definable values that enable
adjustment (without affecting the factory calibration) of the analyzer reading
in the field [see
“Adjusting Analyzer Reading to Match Specific
on page 2-35]. Enter
0
to disable and
1
to enable RATA.
RATA Multiplier
The
RATA Multiplier
parameter is a user definable value that enables
adjustment (without affecting the factory calibration) of the analyzer response
(or slope) in the field [see
“Adjusting Analyzer Reading to Match Specific
on page 2-35]. Enter the required value.
RATA Offset
The
RATA Offset
parameter is a user definable value that enables adjustment
(without affecting the factory calibration) of the analyzer offset in the field [see
“Adjusting Analyzer Reading to Match Specific Standard(s)”
2-35]. Enter the required value.
Set Time - Day
The
Set Time - Day
parameter sets the current day for the clock driving daily
validations. Enter the required value.
<SET PARAMETER MODE>
RATA
0
0:Disable 1:Enable
<SET PARAMETER MODE>
RATA Multiplier
1.00000
Enter a value
<SET PARAMETER MODE>
RATA Offset
0.00000
Enter a value
<SET PARAMETER MODE>
Set Time - Day
07
Enter a value (DD)
Summary of Contents for SpectraSensors SS2000e
Page 2: ......
Page 4: ......
Page 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 113: ......
Page 114: ...www spectrasensors com contact 4900002262 Rev B...