Select Yes to stop the measurement in order to display the SENSOR TYPE AND
SENSOR CONSTANTS window. Refer to Figure 4–33.
Use the arrow buttons to navigate between input fields.
To enter editing mode
1.
Press
to edit the highlighted field.
2.
Change the setting or value (one digit at a time) by pressing the
buttons.
3.
Make the desired change to an input field.
4.
Press
again to save the changes.
To exit editing mode
1.
Press
to cancel and exit.
Changing the sensor type
If it is necessary to change the probe type in the field, change the sensor type
(OP-3, OP-6 or OP-9) according to the sensor that is connected to the analyzer.
The displayed sensor constants (dKSV1, dKSV2, dPhi1, dPhi2, f1 and m) will
change according to the sensor type selected.
The sensor constant values can also be located on the Calibration
Certificate delivered with the optical oxygen sensor. Refer to the
example in Figure 4–34.