OXY5500 Optical Oxygen Analyzer
BA02195C
4
–
16
Hauser
•
Measurement Name
: This box displays the selected measurement
file in which all data is stored when logging is turned on.
•
General
: Here the
Sensor type
of the currently connected oxygen
sensor is displayed.
– The currently measured or manually set
Pressure
value is
also displayed in the General box. For automatic measurement,
the display will show the interpreted pressure value from the
4-20mA input
.
– In the lower right of the
General
box, the time
Interval
at which
the measurements are taken is displayed.
–
Next
provides the time period (the countdown during a running
measurement) until the next measurement.
– The
RATA
is displayed at the bottom of the screen.
–
Error Codes
are also displayed under the
General
box. Refer to
Table 4–1. Error codes are also logged together with
measurement data. During error-free measurements, a 0 value
will display.
•
Press
to return to Simple view.
•
Press
to display the measurement graph. Pressing the button
will switch to the graphical presentation of the current
17.
•
Press
to return to the MAIN MENU screen.
Error codes
The error code is a bit combination of multiple errors. Table 4–1 shows a list of
error bits. For more information on error notifications, refer to Table B–3 on
page B–18. Some examples of error codes are shown below:
“Error code:
1” = No RTD (Pt100) (bit O)
The temperature unit can be changed in Manual mode. Values
ranging from -99
°
C to 199
°
C can be input into the MEAS.
The measurement file can be changed in the MEAS. SETTINGS
menu. Refer to
23.
If no pressure sensor is connected, the display will read 1013 mbar.