BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
3
–
7
3.
For AC systems, pull ground, neutral (N) and L1 wires into the
electronics enclosure. Refer to Figure 3–2.
For DC systems, pull VI +, – and ground wires into the electronics
enclosure. Refer to Figure 3–2.
4.
Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal block.
5.
Connect the main ground wire to the protection ground terminal
marked .
6.
Close and tighten the analyzer enclosure door.
An approved switch or circuit breaker rated for 15 amps should be
used and clearly marked as the disconnecting device for the
analyzer.
Apply 2.25 nm (20 in-lbs) of torque on each bolt to ensure the
door is closed properly to maintain required ingress protection.
Figure 3–2
AC/DC power connections
AC POWER CONNECTION
DC POWER CONNECTION
V1 + –
DC LO
DC ON
N