BA02195C
OXY5500 Optical Oxygen Analyzer
Hauser
B
–
7
Required tools
•
Flathead screwdriver (standard size and mini)
•
9/16 in. open-ended wrench
•
Adjustable wrench
•
10 in. crescent wrench
Removing the pressure sensor
1.
Turn off the power to the analyzer and open the enclosure door using
a standard flathead screwdriver to unlatch the lock.
2.
With a 9/16 in. wrench, loosen the Swagelok nut closest to the
pressure sensor.
3.
Using the same wrench, loosen the Swagelok nut on the T-fitting.
Refer to Figure B–6.
Figure B–6
Removing the Swagelok nuts
PRESSURE
SENSOR NUT
T-FITTING NUT