Solitrend MMP40
Installation
Hauser
11
5
Installation
5.1
Installation conditions
• The device must be installed at a point in the process in such a way that ensures a
relatively constant material density, as the material density is directly related to the
measured value/reading. Where necessary, a bypass should be created or structural
measures may be needed at the place of installation to ensure that the material flow,
and therefore the material density, over the sensor is relatively constant.
• The measuring cell should be completely surrounded by material. For this type of device,
the minimum layer of material covering the cell must be ≥35 mm (moisture-dependent).
• The flow of material over the surface of the sensor should be relatively continuous.
Depending on the operating mode, the electronics module offers ways to automatically
detect and bridge gaps in material in intervals of seconds.
• No material deposits or buildup may form on the sensor surface, as this would falsify the
readings.
Longer averaging times increase measurement accuracy.
5.2
Mounting location for round sensor, short / middle
version
The short / medium round sensor can be installed using a mounting flange.
1
2
3
A0037422
2
Round sensor mounted, rear view
1
Vessel wall
2
Mounting flange
3
Round sensor
5.2.1
Mounting flange Ø108 mm
The mounting flange for the round sensor (short or middle version) can be mounted on
the bottom or on the side wall of the container.