Display and Operating Interface
Display and operating elements in IP 66 field housing
Display (LCD)
– 4½-character display, optional lighting
– with segment display of current in 10 % steps
–
Display elements: error, signal overflow or underflow,
communication
LEDs
– one yellow LED each for indicating the switching status
of the relay (lit = relay energised)
– one yellow LED error indication (lit = error-free operation)
– one green LED indicates error-free indication
(lit = error-free operation, flashing = warning)
Counter
– FMU 861 standard only, FMU 862 optional
– Version: six-character, cannot be reset
Software counter
– FMU 861 standard only, FMU 862 optional
Communication interfaces
Handheld terminals
– INTENSOR: Commulog VU 260 Z
–
HART: Universal HART-Communicator DXR 275
Commuwin II operating program from Hauser
– via Commubox FXA connection to serial interface
RS 232C of a PC
Rackbus RS 485
– via FXA 675 interface board connection to Rackbus
RS 485 network
Synchronising connection
– Parallel link for up to 20 instruments if several
sensor cables are laid together over long distances.
Power Supply
AC
180…253 V (50/60 Hz), 90…132 V (50/60 Hz),
38…55 V (50/60 Hz), 19…28 V (50/60 Hz)
DC
Power consumption
20…30 V (residual ripple within tolerances)
max. 12 W (typical 8 W), max. 500 mA at 24 V DC
Ripple with Smart transmitters
INTENSOR max. ripple (measured to 500 W) 0…500 kHz:
U
PP
=30 mV
HART max. ripple (measured to 500 W) 47…125 Hz:
U
PP
=200 mV
max. noise (measured at 500 W) 500 Hz…10 kHz:
U
eff
=2.2 mV
Reliable galvanic isolation
between current output, relay outputs, interfaces, power
supply and sensor inputs
Sensor cable
common, two-wire screened cable
Maximum values: to 6
Ω
per wire, max. 60 nF
Certificates and Approvals
Ignition protection
see Product Structure Page 15
CE Mark
By attaching the CE Mark,, Hauser confirms that
the instrument fulfils all the requirements of the relevant EC
directives.
Order Code
See "Product Structure"
Supplementary Documentation
Technical Information TI 189F/00/en Prosonic Ultrasonic Sensors FDU 80…86
Operating Instructions BA 100F/00/en Prosonic FMU 860…862
14