3 Installation
Proline Prosonic Flow 90
16
Hauser
3.2.5
Inlet and outlet runs (Insertion version)
If possible, install the sensor well clear of assemblies such as valves, T-pieces, elbows, etc. The long-
est inlet or outlet run must always be taken into account if several flow obstructions are built in.
Compliance with the following requirements for the inlet and outlet runs is recommended to ensure
measuring accuracy.
A0001107
Fig. 9:
Inlet and outlet runs (Insertion version)
1 = Valve; 2 = Pump; 3 = Two pipe bends in different directions
Data above the dimension line: apply to the single path version
Data below the dimension line: apply to the dual path version
3.2.6
Connecting cable length
Shielded cables are offered in the following lengths:
5 m, 10 m, 15 m and 30 m
"
Caution!
Route the cable well clear of electrical machines and switching elements.
1
2
3
³
15 x DN
³
10 x DN
³
5 x DN
³
5 x DN
³
40 x DN
³
40 x DN
³
40 x DN
³
20 x DN
³
20 x DN
³
15 x DN
Summary of Contents for Prosonic Flow 90
Page 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Page 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Page 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Page 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Page 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Page 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Page 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Page 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Page 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...