Proservo NMS83
Operability
Hauser
53
7.3.2
Standard view (measured value display)
X X X X
X X X
X
X
5
2
8
1
3
6
mm
4841.00
Level bal.
F
4
7
A0028702
25
Typical appearance of the standard view (measured value display)
1
Display module
2
Device tag
3
Status area
4
Display area for measured values
5
Display area for status symbols
6
W&M indication
7
Gauge status indication
8
Balance flag; displacer movement
Status symbols
Symbol
Meaning
A0013956
"Failure"
A device error is present. The measured value is no longer valid.
A0013959
"Function check"
The device is in service mode (e.g. during a simulation).
A0013958
"Out of specification"
The device is operated:
• Outside of its technical specifications (e.g. during startup or a cleaning)
• Outside of the configuration carried out by the user (e.g. level outside configured span)
A0013957
"Maintenance required"
Maintenance is required. The measured value is still valid.
Measured value symbols
Symbol 1
Symbol 2
Measured value
A0028148
• Tank level
• Measured level
• Tank level %
A0028149
Water level
A0028528
Liquid temperature
A0028528
A0027990
Vapor temperature
A0028528
A0027991
Air temperature
A0027993
• Tank ullage
• Tank ullage %
A0028150
Observed density value
Summary of Contents for Proservo NMS83
Page 317: ......