Operation
Proline Prosonic Flow 91
44
Hauser
4.3.5
Device status/diagnosis messages
You can read the extended device status, in this case, current diagnosis messages, via
Command "48". The command delivers bit-encoded information (see table below).
!
Note!
• For detailed information on the device status/diagnosis messages and how they are rectified
• Bits and bytes not listed are not assigned.
Byte
Bit
Diagnosis code
Brief description of the diagnosis message
0
7
C - 284
Software update
6
C - 481
Diagnosis active
5
C - 281
Initialization
4
C - 411
Upload/download
3
F - 001
Device fault
2
F - 282
Data storage
1
F - 283
Memory contents
0
F - 062
Sensor connection - Down
1
7
F - 062
Sensor connection - Up
6
F - 881
Sensor signal
5
C - 431
Calibration
4
C - 412
Writing backup
3
C - 413
Reading backup
2
C - 461
Signal output - Current adjust
1
C - 453
Hide value - Pos. Zero return
0
C - 484
Simulation error
2
7
C - 485
Simulation value
6
C - 482
Simulation output - Current
5
C - 482
Simulation output - Frequency
4
C - 482
Simulation output - Pulse
3
C - 482
Simulation output - Status
2
S - 461
Signal output -Current
1
S - 461
Signal output - Frequency
0
S - 461
Signal output - Pulse
3
0
S - 437
Configuration - Sound velocity
1
S - 437
Configuration - Interference
2
-
-
3
-
-
4
-
-
5
-
-
6
-
-
7
-
-