Installation
Proline Promass
24
Hauser
RUPTURE DISK
i
1
DN 8 ( ")...
( ")
!
⁄#
50 2
DN
( ")
80 3
A0029956
9
Rupture disk label
PromassG
Make sure that the function and operation of the rupture disk is not impeded through the
installation of the device. The position of the rupture disk is indicated on a sticker applied over
it. If the rupture disk is triggered, the sticker is destroyed. The disk can therefore be visually
monitored.
RUPTURE DISK
i
1
A0030005
Promass X
Make sure that the function and operation of the rupture disk is not impeded through the
installation of the device. The position of the rupture disk is indicated on a sticker beside it.
The transportation guard must be removed.
The existing connecting nozzles are not intended for the purpose of rinsing or pressure
monitoring, but instead serve as the mounting location for the rupture disk.
In the event of a failure of the rupture disk, a discharge device can be screwed onto the
internal thread of the rupture disk in order to drain off any escaping medium.
Summary of Contents for Proline Promass series
Page 32: ...www addresses endress com...