Proline Promass Q 300
10
Hauser
Recommended measuring range
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value do not override the electronics unit, with the result that
the totalizer values are registered correctly.
Input signal
Input and output versions
External measured values
Various pressure transmitters and temperature measuring devices can be ordered from Endress
+Hauser: see "Accessories" section → 83
It is recommended to read in external measured values to calculate the following measured variables
for gases:
• Mass flow
• Corrected volume flow
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Current input
The measured values are written from the automation system to the measuring device via the
Digital communication
The measured values can be written from the automation system to the measuring via:
• FOUNDATION Fieldbus
• PROFIBUS PA
• Modbus RS485
Current input 0/4 to 20 mA
Current input
0/4 to 20 mA (active/passive)
Current span
• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Resolution
1 µA
Voltage drop
Typically: 0.6 to 2 V for 3.6 to 22 mA (passive)
Maximum input voltage
≤ 30 V (passive)
Open-circuit voltage
≤ 28.8 V (active)
Possible input variables
• Pressure
• Temperature
• Density
Status input
Maximum input values
• DC –3 to 30 V
• If status input is active (ON): R
i
>3 kΩ
Response time
Adjustable: 5 to 200 ms