Proline Promass Q 300
42
Hauser
RUPTURE DISK
i
1
2
3
A0030346
1
Rupture disk label
2
Rupture disk with 1/2" NPT internal thread with 1" width across flat
3
Transport protection
For information on the dimensions: see the "Mechanical construction -> Accessories" section
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration
takes place under reference conditions → 36. Therefore, a zero point adjustment in the field is
generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or very high-
viscosity fluids).
Protective cover
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)
255 (10.0)
134 (5.3)
30 (1.18)
A0029553
Determining the pitch angle and roll angle
For correct measurement, the pitch angle and roll angle must be determined and entered .
The angle can be entered with an accuracy of ±10 °.