Technical data
Proline Promass F 100 Modbus RS485
106
16.9 Process
Medium temperature range
Standard version
–50 to +150 °C (–58 to +302 °F)
Order code for "Measuring tube
mat., wetted surface", option
HA,
SA, SB, SC
Extended temperature version
–50 to +240 °C (–58 to +464 °F)
Order code for "Measuring tube
mat., wetted surface", option
SD,
SE, SF, TH
Dependency of ambient temperature on medium temperature
T
a
T
m
B
A
A0031121
17
Exemplary representation, values in the table below.
T
a
Ambient temperature
T
m
Medium temperature
A
Maximum permitted medium temperature T
m
at T
a max
= 60 °C (140 °F); higher medium temperatures T
m
require a reduced ambient temperature T
a
B
Maximum permitted ambient temperature T
a
for the maximum specified medium temperature T
m
of the
sensor
Values for devices used in the hazardous area:
Separate Ex documentation (XA) for the device .
Not insulated
Insulated
A
B
A
B
Version
T
a
T
m
T
a
T
m
T
a
T
m
T
a
T
m
Standard version
60 °C (140 °F) 150 °C (302 °F)
–
–
60 °C (140 °F) 110 °C (230 °F) 55 °C (131 °F) 150 °C (302 °F)
Extended
temperature
version
60 °C (140 °F) 170 °C (338 °F) 55 °C (131 °F) 240 °C (464 °F) 60 °C (140 °F) 110 °C (230 °F) 50 °C (122 °F) 240 °C (464 °F)
Density
0 to 5
000 kg/m
3
(0 to 312 lb/cf)