Commissioning
Proline Promag W 400 HART
100
Hauser
10.4.9 Configuring empty pipe detection
The
Empty pipe detection
wizard guides you systematically through all the parameters
that have to be set for configuring empty pipe detection.
Navigation
"Setup" menu → Empty pipe detection
‣
Empty pipe detection
Empty pipe detection
New adjustment
Progress
Switch point empty pipe detection
Response time empty pipe detection
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Empty pipe detection
–
Switch empty pipe detection on
and off.
• Off
• On
Off
New adjustment
The
On
option is selected in
the
Empty pipe detection
parameter.
Select type of adjustment.
• Cancel
• Empty pipe adjust
• Full pipe adjust
Cancel
Progress
The
On
option is selected in
the
Empty pipe detection
parameter.
Shows the progress.
• Ok
• Busy
• Not ok
–
Switch point empty pipe detection
The
On
option is selected in
the
Empty pipe detection
parameter.
Enter hysteresis in %, below
this value the measuring tube
will detected as empty.
0 to 100 %
50 %
Response time empty pipe detection In the
Empty pipe detection
On
option is selected.
Enter the time before
diagnostic message S862 ''Pipe
empty'' is displayed for empty
pipe detection.
0 to 100 s
1 s