Proline Promag L 400 HART
Commissioning
Hauser
71
Sequence of the wizard for the pulse output
Failure mode
End of wizard
Invert outp.sig.
Operating mode
Pulse
Volume flow
Assign pulse
Mass flow
Off
Mass unit
Density unit
Fixed density
Volume unit
Pulse width
Value per pulse
A0017435-EN
å
23
"Pulse/frequency/switch output 1-2" wizard in the "Setup" menu: "Pulse" operating mode
Parameter overview with brief description for pulse output
Parameter
Description
Selection/
User entry
Factory setting
Operating mode
Specify the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
The picklist depends
on the selected device
order (
Pulse
Assign pulse output
Select the process variable for
the pulse output.
• Off
• Volume flow
• Mass flow
Off
Volume unit
Select the unit for volume.
Result
The selected unit is taken
from:
Volume flow unit
Unit choose list
Depends on country and
nominal diameter