Commissioning
Proline Promag L 400 HART
68
Hauser
Sequence of the wizard
Reset totaliz.
1…3
Assign stat.inp.
Off
End of wizard
Active level
Reset all tot.
High
Low
Flow override
Response time
A0017422-EN
å
21
"Status input" wizard in the "Setup" menu
Parameter overview with brief description
Parameter
Description
Selection/
User entry
Factory setting
Assign status input
Select the function for the
status input.
• Off
• Reset totalizer 1
• Reset totalizer 2
• Reset totalizer 3
• Reset all totalizers
• Flow override
Off
Active level
Specify the input signal level
at which the assigned
function is triggered.
• High
• Low
High
Response time status input
Specify the minimum
amount of time the input
signal level must be present
before the selected function
is triggered.
5 to 200 ms
50 ms