Installation
Proline Promag L 400 HART
30
Hauser
6.3
Post-installation check
Is the device undamaged (visual inspection)?
q
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature
• Process pressure (refer to the section on "Pressure-temperature ratings" in the "Technical Information"
document)
• Ambient temperature range
• Measuring range
q
Has the correct orientation for the sensor been selected ?
• According to sensor type
• According to medium temperature
• According to medium properties (outgassing, with entrained solids)
q
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the piping ?
q
Are the measuring point identification and labeling correct (visual inspection)?
q
Is the device adequately protected from precipitation and direct sunlight?
q
Have the fixing screws been tightened with the correct tightening torque?
q