Proline Promag 50H, 53H
Hauser
17
Length of connecting cable
When mounting the remote version, please note the following to achieve correct measuring results:
• Fix cable run or lay in armored conduit. Cable movements can falsify the measuring signal especially in the
case of low fluid conductivities.
• Route the cable well clear of electrical machines and switching elements.
• If necessary, ensure potential equalization between sensor and transmitter.
• The permitted cable length L
max
is determined by the fluid conductivity. A minimum conductivity of
20
μ
S/cm is required for measuring demineralized water.
• When the empty pipe detection function is switched on (EPD),
the maximum connecting cable length is 10 m (33 ft).
A0010734
Permitted length of connecting cable for remote version
Area marked in gray = permitted range; L
max
= length of connecting cable in [m] ([ft]); fluid conductivity in [
μ
S/cm]
200
100
5
10
100
200
L
max
[m]
[m]
[µS/cm]
L
max
[ft]
200
600
0
400