Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
99
3.4
"Input" submenu
Navigation
Expert → Input
‣
Input
‣
Current input 1 to n
‣
Status input 1 to n
3.4.1
"Current input 1 to n" submenu
Navigation
Expert → Input → Current input 1 to n
‣
Current input 1 to n
Terminal no. (1611–1 to n)
Signal mode (1610–1 to n)
Current span (1605–1 to n)
0/4 mA value (1606–1 to n)
20 mA value (1607–1 to n)
Failure mode (1601–1 to n)
Failure value (1602–1 to n)
Terminal no.
Navigation
Expert → Input → Current input 1 to n → Terminal no. (1611–1 to n)
Description
Displays the terminal numbers used by the current input module.
User interface
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
Additional information
"Not used" option
The current input module does not use any terminal numbers.