11 Description of Device Functions
PROline Promag 10
110
Hauser
POSITIVE ZERO
RETURN
Use this function to interrupt evaluation of measured variables. This is
necessary when a piping system is being cleaned, for example.
This setting acts on all function and outputs of the measuring device.
Options:
OFF
ON
→
Signal output is set to the “ZERO FLOW” value.
Factory setting:
OFF
SYSTEM DAMPING
Use this function to set the filter depth of the digital filter. This reduces the
sensitivity of the measuring signal to interference peaks (e.g. high solids
content, gas bubbles in the fluid, etc.). The system reaction time increases
with the filter setting.
User input:
0 ... 4
Factory setting:
3
!
Note!
The system damping acts on all functions and outputs of the measuring
device.
Function description SYSTEM PARAMETER