Device functions - Coriolis measuring systems
Proline Fieldcheck
72
Hauser
Group VERIFICATION SENSOR
Function description
FUNCTION
VERIFICATION SENSOR
INSTRUMENT
VERSION
Use this function to enter the measuring device version.
Options:
COMPACT / REMOTE:
Remote version with a cable length of up to 20 meters max. or compact version
REMOTE > 20 m:
Remote version with a cable length longer than 20 meters
!
Note!
The HV isolation test is not available for the remote version with a cable length longer
than 20 meters.
Factory setting:
COMPACT / REMOTE:
ISOLATION TEST
Use this function to select whether an extended isolation test (against undesired ground
connections) should be carried out for the excitation and sensor coils and the
temperature sensors.
Options:
WITH HV TEST PULSE
NO HV TEST PULSE
Factory setting:
WITH HV TEST PULSE
APPLICATION
REFERENCE DATA
Use this function to select whether the reference data of the application should be
accepted at the end of the sensor verification process.
Options:
YES
NO
Factory setting:
YES
OPERATION
Use this function to start the sensor verification, cancel a verification that is running or
save data after verification has been completed.
Options:
START
STOP
STORE DATA?
Factory setting:
START
NEXT ACTION
You are prompted to select the next step once the sensor verification has ended.
Options:
Simulation/verification
Fieldcheck end