Proline Fieldcheck
Device functions - Coriolis measuring systems
Hauser
63
APPLIC. SPEC. DATA 1
TEMP M/C
Use this function to select the application-specific temperature for the carrier and
measuring tube in the simulated process.
Options:
–25, +15, +60, +125 °Celsius
–10, +60, +135, +255 °Fahrenheit
+250, +290, +335, +400 Kelvin
+450, +520, +600, +715 Rankine
APPLIC. SPEC. DATA 2
DENSITY
Use this function to enter the application-specific density in the simulated process.
Options:
0.400 to 2.000 kg/l
or the corresponding values in the unit used in the measuring device.
SIGNAL 1 LEVEL/
SIGNAL 2 LEVEL
Use this function to enter the flow value which is to be simulated.
Enter the second flow value which is to be simulated in SIGNAL 2 LEVEL if you have
selected the STEP operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
0.000 to 999999 (6-digit floating-point number with sign)
!
Note!
The flow velocity must lie within the value range -12.5 m/s to +12.5 m/s.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
0.000
SIGNAL 1 PERIOD/
SIGNAL 2 PERIOD
Use this function to enter the duration for simulation signals 1 and 2 if you have selected
the STEP operating mode.
!
Note!
More information is provided in the OPERATION MODE function.
User input:
1 to 999 s (seconds)
Factory setting:
10 s
Function description
FUNCTION
SIMULATION
CONFIGURATION