Wiring
Proline Fieldcheck
20
Hauser
3.6.1
Connecting sensor test box to Promass 80, 83, 84 wall-mount
housing
!
Note!
The start-up procedure must have been executed via the Simubox before connecting the sensor test
box.
"
Caution!
Connect the ground terminals of the adapter to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Connect the sensor test box to Fieldcheck with the simulation cable.
2.
Connect the sensor test box to the sensor cable connector and the cable ends of the excitation
coil connector with the Coriolis adapter.
A0013405
Fig. 13:
Coriolis adapter for sensor test box
1
Sensor test box
2
Coriolis adapter
3
Sensor cable connector
4
Excitation current cable
5
Ground terminal
6
Connection to Fieldcheck (with Simubox cable)
2
3
4
5
1
6