6.3
Simulated operating mode
This function is intended primarily for checking the correct function of the system:
•
Enter 6 at V8H0 to activate the simulation mode for channel 1
•
Enter 7 at V8H0 to activate the simulation mode for channel 2
•
Enter 0 at V8H0 to terminate simulation and resume normal measurements.
When the simulation mode is activated, the alarm relay flashes (Warning E613 or E 614).
The following simulations are possible:
Matrix
Entry
Simulated variable
V9H6
Frequency (Hz)
Frequency, level, volume, current
V9H7
Level
Level, volume, current
V9H8
Volume
Volume, current
V9H9
Current (-2
…
+22 mA)
Current
The level simulation mode takes the last measured value as default value in V9H7.
Example:
Simulation of volume
and current on channel 1
by entry of level in V9H7
#
Matrix
Entry
Significance
1
V8H0
6
Simulation Ch. 1
2
-
»E«
Register entry
3
V9H7
e.g. 80% Enter level
4
-
»E«
Register entry
5
V9H8
**.**
Volume for level
6
V9H9
**.**
Current for level
7
V8H0
e.g. 0
Measurement mode
8
-
»E«
Register entry
Example:
Simulation of current on
channel 2 by entry of
volume in V9H8
#
Matrix
Entry
Significance
1
V8H0
7
Simulation Ch. 2
2
-
»E«
Register entry
3
V9H8
e.g. 500
Volume = 500 hl
4
V9H9
**.**
Current for volume
5
V8H0
e.g. 2
Measurement mode
6
-
»E«
Register entry
Example:
Simulation of current on
channel 1 by entry of
current in V9H9
#
Matrix
Entry
Significance
1
V8H0
6
Simulation Ch. 1
2
-
»E«
Register entry
3
V9H9
16
16 mA
4
V8H0
e.g. 0
Measurement mode
5
-
»E«
Register entry
V9H7
le
vel
uni
ts
volume, current mA
V9H8
vol
u
m
e
uni
ts
current mA
BA143_42
Fig. 6.2:
Simulation mode
Chapter 6: Trouble-Shooting
Prolevel FMC 662
40