Installation
10
Hauser
3.4
Installation examples
3.4.1
Fixed installation (CPA442)
The CPA442 fixed installation assembly allows a sensor to be easily adapted to almost any process
connections from Ingold nozzles to Varivent or Triclamp connections.
This type of installation is very suitable for tanks and larger pipes. The sensor achieves a defined
immersion depth in the medium in an easy manner.
3.4.2
Flow assembly
CPA240
The CPA240 flow assembly offers up to three installation slots for sensors with a shaft diameter of 12
mm (0.47"), a shaft length of 120 mm (4.72") and a process connection Pg 13.5.
It is very suitable for use in pipes or at hose connections.
During trace measurement in particular, ensure the assembly is completely vented to avoid measuring
errors.
a0005720
Fig. 3: CPA240 flow assembly with protective hood
a0005721
Fig. 4: Bypass
installation
1
Main line
2
Medium removal
3, 6
Manually actuated or solenoid valves
4
Sampling
5
Flow assembly with installed sensor
7
Medium return