Maintenance
OUSAF11
20
Hauser
5.
Remove the lamp from the fitting and replace it with a new one. Use a paper towel to
insert the new lamp.
6.
Without graphics:
Once you have inserted the new lamp, clean it with a paper towel and ethanol, if
available. Make sure you do not touch the optical surfaces again after cleaning.
7. Insert the lamp module back into the sensor head until it is firmly seated in the correct
position. Tighten the securing screw slightly.
8. Mount the sensor again. In doing so, use PTFE tape to seal the thread of the sensor
head.
Check the function of the sensor afterwards.
Always replace the O-ring each time you replace the lamp. Read the following section for this
purpose.
8.4
Replacing the O-ring
‣
Switch off the lamp at the transmitter using the software function.
‣
Allow the lamp to cool down (30 minutes).
1.
Release the cable gland and pull it over the cable to remove it from the sensor.
Summary of Contents for OUSAF11