Maintenance
OUSAF11
22
Hauser
4.
If necessary, sever the cable tie on the sensor cable. There is a new cable tie in the
spare parts kit which you can use to ensure strain relief on the cable once you re-insert
the lamp and detector unit.
Remove the lamp and detector unit from the sensor head.
Sensor versions with Tri-Clamp or VARIVENT have one O-ring on either side of the
process connection. Always replace the two O-rings.
5.
a
Release the O-ring (item a) from the guide in the sensor shaft.
6.
Pull the O-ring over the lamp and detector unit and dispose of it correctly.
7.
Without graphics:
Push the new O-ring over the lamp and detector unit into the guide on the sensor shaft.
8. Re-install the lamp and detector unit. Tighten the securing screw slightly.
9. Mount the sensor again. In doing so, use PTFE tape to seal the thread of the sensor
head.
Check the function of the sensor afterwards.