Model OUM980 with OUSAF21 Inline Sensor
Commissioning
36
The next screen is for entry of target correlation values that can be
assigned to the original correlated measured value. i.e. 1.32 measured
units to 9999.9 units. The FROM value is the last entered value on the
‘ENTER USER VALUE’ screen. Enter the new value using
and
to position the cursor and select the
number for each required position. If a different position is required for
the decimal point, do this first. Press
twice to change and
accept.
Repeat above procedure until all correlation points have been entered.
When all the sample points have been edited, the Change Units screen
is next. The engineering unit can be entered is limited to 6 alpha-
numeric characters.( i.e. gr/l, %, mg/l, cel/cc, etc.). Press
twice to change and accept entry.
The Max/Min values are calculated for entered values and units.
These values are the new limits for output span and relay setting and
the displayed Process Display PV value and units.
Note!
If the OD to correlation unit table is predefined and does not need liquid sampling, use MANUAL EDIT option to entry the table to
analyzer. The procedure is similar to SAMPLE EDIT.
Switch Correlation Curves
Enter PROCESS SET-UP sub-menu to access RUN CAL CURVE
function. Press
to continue.
The next screen asks which curve (0-3) you want to execute. Curve
#0 is the default OD measurement. Curves 1-3 are user defined. Scroll
with
to select curve number. Press
to select and
confirm.
NOTE:
Entering a curve with Process Correlate will automatically set it as the
current active curve.