Repair
13.1 General information
13.1.1 Repair concept
Hauser repair concept
• The devices have a modular design
• Repairs can be carried out by Hauser Service or by appropriately trained
customers
• Spare parts are grouped into logical kits with the associated replacement instructions
For more information on service and spare parts, please contact your Hauser
sales representative.
13.1.2 Repair of Ex-certified devices
L
WARNING
Incorrect repair can affect electrical safety!
Explosion Hazard!
‣
Only specialist personnel or the Hauser service team may carry out repairs on
Ex-certified devices.
‣
Relevant standards and national regulations on hazardous areas, safety instructions
and certificates must be observed.
‣
Use only original Hauser spare parts.
‣
Please note the device designation on the nameplate. Only identical parts may be used
as replacements.
‣
Carry out repairs according to the instructions.
‣
Only the Hauser service team is permitted to modify a certified device and
convert it to another certified version.
‣
All repairs and modifications must be documented.
13.2 Spare parts
• Some replaceable device components are identified by a spare part nameplate. This
contains information about the spare part.
• All the spare parts available for the measuring device are listed in the
Device Viewer
(
) along with the order code and can be ordered. If
available, users can also download the associated Installation Instructions.
Device serial number or QR code:
Located on the device and spare part nameplate.
13.3 Replacement
L
CAUTION
Data upload/download is not permitted if the device is used for safety-related
applications.
‣
After an entire device or an electronics module has been replaced, the parameters can
be downloaded to the device again via the communication interface. For this, the data
must have been uploaded to the PC beforehand using the "FieldCare/DeviceCare"
software.