Diagnostics and troubleshooting
Micropilot FMR51, FMR52 HART
74
Hauser
12
Diagnostics and troubleshooting
12.1 General trouble shooting
12.1.1 General errors
Error
Possible cause
Remedial action
Device does not respond.
Supply voltage does not match the
value indicated on the nameplate.
Connect the correct voltage.
The polarity of the supply voltage is
wrong.
Correct the polarity.
The cables do not contact the
terminals properly.
Ensure electrical contact between the
cable and the terminal.
Values on the display
invisible
Contrast setting is too weak or too
strong.
• Increase contrast by pressing and
simultaneously.
• Decrease contrast by pressing and
simultaneously.
The plug of the display cable is not
connected correctly.
Connect the plug correctly.
Display is defective.
Replace display.
"Communication error" is
indicated on the display when
starting the device or
connecting the display
Electromagnetic interference
Check grounding of the device.
Broken display cable or display plug.
Exchange display.
Output current < 3.6 mA
Signal cable connection incorrect.
Check connection.
Electronics is defective.
Replace electronics.
HART communication does
not function.
Communication resistor missing or
incorrectly installed.
Install the communication resistor (
250 Ω) correctly → 35.
Commubox connected incorrectly.
Connect Commubox correctly → 48.
Commubox not switched to HART
mode.
Set the selection switch of the
Commubox to the HART position.
CDI communication does not
work.
Wrong setting of the COM port on
the computer.
Check the setting of the COM port on
the computer and change it if necessary.
Device measures incorrectly.
Parametrization error
Check and adjust parameterization.
12.1.2 Parametrization errors
Error
Possible cause
Remdy
Measured value incorrect
If measured distance (Setup
→ Distance) matches the real
distance:
Calibration error
• Check and adjust
Empty calibration
parameter
• Check and adjust
Full calibration
parameter (→ 109) if necessary.
• Check and adjust linearization if
necessary (
Linearization
submenu
For measurements in bypasses /
stilling well:
• Wrong tank type
• Wrong tube diameter
• Select
=
Bypass / pipe
.
• Enter correct diamter in
Tube
diameter
Enter correct diamter in
Enter correct value in
Level correction
parameter (→ 122).