Installation
Micropilot FMR51, FMR52 HART
24
Hauser
100%
0%
B
A
C
H
øD
A0018872
Device
A [mm (in)]
B [m (ft)]
C [mm (in)]
H [m (ft)]
FMR51
50(1.97)
> 0.2 (0.7)
50 to 250 (1.97 to 9.84)
> 0.3 (1.0)
FMR52
200(7.87)
6.3
Mounting cladded flanges
• Use flange screws according to the number of flange holes.
• Tighten the screws with the required torque (see table).
• Retighten the screws after 24 hours or after the first temperature cycle.
• Depending on process pressure and process temperature check and retighten the
screws at regular intervals.
Usually, the PTFE flange cladding also serves as a seal between the nozzle and the
device flange.
Flange size
Number of screws
Recommended torque [Nm]
minimum
maximum
EN
DN50/PN16
4
45
65
DN80/PN16
8
40
55
DN100/PN16
8
40
60
DN150/PN16
8
75
115
ASME
2"/150lbs
4
40
55
3"/150lbs
4
65
95
4"/150lbs
8
45
70
6"/150lbs
8
85
125