Micropilot FMR10
Operating menu
Hauser
59
Selection
• Off
• 4 mA
• 20 mA
• Calculate
• Reset
Factory setting
Off
Trim value high
Navigation
Setup → Advanced setup → Current output → Trim value high
Description
Enter upper measured value for the trim (around 20 mA). After this value has been
entered: Select 'Trim' = Calculate. This initiates the recalibration of the current output.
User entry
18.0 to 22.0 mA
Factory setting
20.0 mA
Trim value low
Navigation
Setup → Advanced setup → Current output → Trim value low
Description
Enter lower measured value for the trim (around 4 mA). After this value has been entered:
Select 'Trim' = Calculate. This initiates the recalibration of the current output.
User entry
3.0 to 5.0 mA
Factory setting
4.0 mA