5.2.2
Setup 2
Coding
Field
Selection or range
default (bold)
Display
Info
A3
Enter altitude
0 m
0 ... 4000 m
Only available if the pressure
compensation in A2 is switched
off or not available.
A4
Enter measured
value damping
1
1 ... 60
Measured value damping causes
averaging over the specified
number of individual measured
values.
This is used for example, to
stabilise the display if the
measurement is unstable.
There is no loss if “1" is entered.
A5
Enter oxygen
measuring
range
20 mg/l
60 mg/l
200 % Sat
600 % Sat
400 hPa
1200 hPa
Sensor COS 4 / 4HD / 41 / 71
:
Measuring range must be
0-20 mg/l (0-200 % Sat,
0-400 hPa).
Sensor COS 3 / 3S / 3HD / 31:
Both measuring ranges possible.
The factory settings are printed in
bold
; functions printed in
italics are not available in the basic version.
Coding
Field
Selection or range
Factory setting
Display
Info
B
Function group
SETUP 2
Initial display in function group
SETUP 2.
B1
Salinity
0.0 %
0.0 ... 4.0 %
B2
Enter correct
process
temperature
current meas. value
–10 ... 60.0 °C
The display value can be edited.
You can make an adjustment of
max. ± 5 °C. Due to the high
accuracy, adjustment is not
usually necessary.
B3
Displays
temperature
difference (offset)
Current offset
5.0 ... 5.0 C
The offset is the difference
between measured and entered
temperature.
0
A3
Altitude
m
1
A4
Damping
20
A5
Range
%
mg/l
B
SETUP 2
0.0
B1
Salinity
%
0.0
B2
RealTemp
°C
0.0
B3
Temp.Offs
°C
Instrument configuration
Liquisys M COM 223 / 253
26