Error and Status Messages
Error Messages
Error
Message
Remedy
blinks
Replace the batteries
Replace the batteries.
ERR 1
Primary variable range
Check whether the measurement conditions
correspond to the adjusted measuring range.
ERR 2
ORP range
ERR 3
Temperature range
ERR 4
Zero point
Thoroughly rinse the sensor and recalibrate.
If this does not help, replace the sensor.
ERR 5
Slope
ERR 6
Cell constant too high/
low
Enter nominal cell constant or calibrate the
sensor using a known solution.
ERR 7
Air pressure range
Check if the opening for the pressure sensor
located on the back of the device is blocked.
ERR 8
Identical buffers!
Use a buffer solution with a different nominal
value before starting the next calibration step.
ERR 10
Buffers interchanged!
Repeat calibration.
ERR 11
Unstable value
(Drift too high)
Leave the sensor in the liquid until the
measured value is stable. If this does not help,
replace the sensor.
ERR 14
Time and date invalid
Set the date and time.
ERR 18
System error
Restart, reset to factory settings, configure and
calibrate. If the error occurs again, contact the
Service.
ERR 19
Factory settings error
Data error, measurement with analog sensors
no longer possible. Contact the Service team.
ERR 21
No sensor connected
Connect operational Memosens sensor.
ERR 30
Data logger full
Clear the logger completely or partially.
ERR 31
MemoLog full
Clear the MemoLog completely or partially.