15

pH

Oxy Cond

Calibration Record

Shows the data of the last calibration performed on the currently connected sensor.

Information

1)  In measuring mode, press the Menu softkey. 
2)  Select “Information” and confirm by pressing E.
3)  Select the desired submenu and confirm by pressing E.  

The different submenus are described below.

Sensor Information (Digital Sensors only)

Shows the data of the currently connected digital sensor. You can save sensor data 
(MemoLog) in the device by pressing the “Save” softkey. The following table shows 
the sensor information depending on the respective sensor type: 

pH

Cond

Oxy

ISFET

ORP

Manufacturer

P

P

P

P

P

Order No.

P

P

P

P

P

Sensor serial no.

P

P

P

P

P

Cap serial no.
SW version

P

P

P

P

P

HW version

P

P

P

P

P

Calibration*

P

P

P

P

P

Zero point

P

P

Slope 

P

P

P

Sensor operating time

P

P

P

P

P

Wear

P

P

P

SIP

P

P

P

P

P

CIP

P

Cell constant

P

Operating point

P

ORP correction

P

* Last calibration

Summary of Contents for Liquiline To Go CYM290

Page 1: ...Operating Instructions Liquiline To Go CYM290 Products Services Solutions ...

Page 2: ......

Page 3: ...f the device has been in contact with process fluids it must be decontaminated disinfected before shipment In that case please attach a corresponding certificate for the health and safety of our service personnel Disposal Please observe the applicable local or national regulations concerning the disposal of waste electrical and electronic equipment Registered Trademarks The following names are reg...

Page 4: ...13 Switching On the Meter 14 Icons 14 Information 15 Calibration Record 15 Sensor Information Digital Sensors only 15 Sensor Network Diagram pH and Oxy only 16 Sensor Monitor 17 Messages 17 MemoLog Memosens only 17 Device Info 18 Device Test 18 Configuration 19 Calibration 26 Manual Calibration 27 Data Entry Calibration 27 Auto Calibration 28 Entry of Solution Calibration 29 Cell Constant Calibrat...

Page 5: ...Data Logger 38 Viewing the Logger Data 38 Deleting the Logger Data 38 Error and Status Messages 41 Info and Help Texts 42 Error Messages 43 pH Product Line 44 pH Sensors 44 Endress Hauser Buffer Solutions pH 45 Ready to use quality pH buffer solutions 45 Accessories for pH 45 Conductivity Product Line 46 Conductivity Sensors 46 Conductivity Calibration Solutions CLY11 47 Accessories for Conductivi...

Page 6: ...ess The package of the Liquiline To Go CYM290 includes Liquiline To Go CYM290 Meter incl 4 batteries AA and premounted quiver P Carrying strap P Data carrier with detailed user manuals P USB cable 1 5 m P Safety instructions P Quickstart instructions in various languages P ...

Page 7: ...s Installation and first steps Operation Menu structure Calibration Error messages and recommended actions Various languages on CD ROM German English French Italian Spanish Portuguese Brazil Specific Test Report CAUTION These safety instructions are part of the product documentation and must be observed In official EU languages and others EU Declarations of Conformity ...

Page 8: ...t Furthermore it can be used for calibration The rugged housing is made of a high performance polymer It provides high impact resistance and dimensional stability even when exposed to extreme moisture Scratch proof clear glass display perfectly readable even after years Long operating time with one set of batteries 4 x AA or use of a Li ion battery for reliable operation even at high or very low o...

Page 9: ...ble and can still be used when the sensor is connected to another Memosens capable device Sensoface Sensoface provides quick information on the sensor condition The three smiley faces as shown on the right represent the sensor condition during measurement and after a calibration When the condition deteriorates a status message gives a hint to the cause Programmed buffers Programmed buffers is a ve...

Page 10: ...r operation Hook A fold out hook on the back allows suspending the meter This leaves your hands free for the actual measurement The rating plate is located beneath the hook Protective Cover and Hook Combined Cover and hook can be joined together to form a benchtop stand allowing com fortable and fatigue free working at a lab bench or desk ...

Page 11: ...the display above the key Arrow keys Selecting Adjusting entries E Confirming an adjustment Switching on off Switching on Immediate access of meas mode Toggling the display Displaying time and date Menu selection Information Return Menu icons Information Calibration Data logger Configuration Softkeys Start logger ...

Page 12: ...the device fails to perform the intended function prolonged storage at temperatures above 70 C severe transport stresses In this case a professional routine test must be performed This test should be carried out at our factory A special lithium ion battery suited to the battery compartment is available for the Liquiline To Go CYM290 The battery is recharged through the USB port A battery icon in t...

Page 13: ...o the meter at a time Separate temperature probe After power on a separate temperature probe is automatically recognized When you want to replace the temperature probe you must switch off the meter and then switch it on again Start Up Memosens sensors have a cable coupling which allows convenient replacement of sensors while the cable remains connected to the meter The connecting cable is connecte...

Page 14: ...s to measuring mode With first a self test is performed and then the calibration data and settings are displayed before the meter switches to measur ing mode Error message Battery charging level Sensor condition Memosens sensor Temperature Icons Important information about the state of the device Left softkey Measured variable Right softkey Start Up ...

Page 15: ...only Shows the data of the currently connected digital sensor You can save sensor data MemoLog in the device by pressing the Save softkey The following table shows the sensor information depending on the respective sensor type pH Cond Oxy ISFET ORP Manufacturer P P P P P Order No P P P P P Sensor serial no P P P P P Cap serial no SW version P P P P P HW version P P P P P Calibration P P P P P Zero...

Page 16: ...r Wear Memosens Parameters which cannot be checked are shown as inactive gray and are set to 100 e g Sensocheck for analog sensors The parameter values should lie between the outer 100 and inner 50 hexa gon When a value enters the inner hexagon 50 the corresponding caption text flashes red see example Information Help Return Network Diagram pH Response time Sensocheck Zero point Calibration timer ...

Page 17: ... must be activated in the configuration menu With MemoLog activated up to 100 calibration records can be directly saved to the device Then the complete Memosens index structure will be recorded after every calibration Information Sensor Monitor Shows the raw values available fom the connected sensor pH analog mV temperature temperature detector temperature resistance pH digital glass mV temperatur...

Page 18: ... listed below A green checkmark shows that the test was successful FLASH program memory FLASH data memory FLASH parameter memory RAM working memory Display test 1 Select Display test and press E 2 The display lights up red green blue and then white 3 Press any key to stop the test Keypad test 1 Select Keypad test and press E 2 Press all nine keys one after the other A green checkmark shows that a ...

Page 19: ...n Cal timer Off On Interval On 00 99 days Cal mode Programmed buffer Manual Data entry Calibration points Auto 1 point 2 point 3 point Buffer set Endress Hauser Mettler Toledo Knick CaliMat Ciba NIST technical NIST standard Hach WTW Hamilton Reagecon DIN 19267 2 00 4 01 6 98 9 95 11 87 2 00 4 01 7 00 9 21 2 00 4 00 7 00 9 00 12 00 2 06 4 00 7 00 10 00 1 68 4 00 7 00 10 01 12 46 1 679 4 006 6 865 9...

Page 20: ...ircular Circular Logger type Shot Interval 00 00 01 12 59 59 00 02 00 Difference 1st difference On Off Delta pH pH 0 0 16 0 pH 1 0 Delta mV 0 2000 mV 1 mV 2nd diff On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval as logger type interval Difference as logger type difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding ...

Page 21: ...Italiano Français Português Auto off Off 5 min 10 min 30 min 60 min Temperature C F Cond sensor Conductivity S cm S m Range selection Auto o ooo µS cm oo oo µS cm ooo o µS cm oooo µS cm oo oo mS cm ooo o mS cm oooo mS cm Calculation Off MΩ cm TC SAL TDS TC compensation TC Linear NLF NaCl HCl NH3 NaOH TC of solution TC 0 20 0 K 1 00 K Ref temp TC 0 100 0 C 25 C 32 212 F 77 F TDS factor TDS 0 9 99 1...

Page 22: ... CLY 11 B CLY 11 C CLY 11 D CLY 11 E Time Date Time format 24 h 12 h Date format dd mm yyyy yyyy mm dd dd mm yyyy mm dd yyyy Time hh mm ss Date Date format as configured Display Appearance Modern b w Lighting Permanent 60 min 30 min 10 min 5 min 1 min 30 sec Brightness Bright Standard Dim indicates that submenus can be opened by pressing E ...

Page 23: ... 0 g kg 1 0 g kg Delta TDS 0 2000 0 mg l 1 mg l 2nd diff On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval as logger type interval Difference as logger type difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Specifications Factory setting Yes No Note Reset to factory settings will also era...

Page 24: ... Calibration timer Off On Interval On 0 99 days MemoLog Off On TAG Off On Time Date Time format 24 h 12 h Date format dd mm yyyy yyyy mm dd dd mm yyyy mm dd yyyy Time hh mm ss Date Date format as configured Oxygen Configuration 1 In measuring mode press the Menu softkey 2 Select Configuration and confirm by pressing E 3 Make the desired adjustments The following table gives you an overview Factory...

Page 25: ...erence 1st difference On Off Delta satura tion 0 200 Air 1 Air Delta conc 0 20 mg l 1 mg l Delta mbar 0 999 99 mbar 1 mbar 2nd diff On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval see logger type interval Difference see logger type difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Speci...

Page 26: ...0 Hach 4 01 7 00 10 01 12 00 WTW 2 00 4 01 7 00 10 00 Hamilton 2 00 4 01 7 00 10 01 12 00 Reagecon 2 00 4 00 7 00 9 00 12 00 DIN 19267 1 09 4 65 6 79 9 23 12 75 2 Immerse the sensor in the 1st 2nd 3rd buffer solution and press Continue repeat this step for each calibration point 3 Finally the calibration data will be displayed You can Apply or Discard these values pH Calibration 1 In measuring mod...

Page 27: ...p for each calibration point Note When using sensors without temperature detector you should adjust the temperature manually before starting calibration see page 32 3 Finally the calibration data will be displayed You can Apply or Discard these values Data Entry Calibration Calibration by entering known sensor values 1 Press the Start softkey 2 Enter the known sensor values for zero and slope 3 Fi...

Page 28: ... 1 136 1 270 1 406 1 542 1 683 1 824 2 114 mS cm CLY 11 D 8 07 9 16 10 29 11 45 12 64 13 86 15 11 16 37 18 97 CLY 11 E 70 58 79 34 88 20 97 56 107 00 116 52 126 10 135 98 155 82 6 Press the Start softkey 7 Immerse the sensor in the solution and press Continue 8 Finally the calibration data record will be displayed You can Apply or Discard these values NOTICE Make sure that the values of the calibr...

Page 29: ...he calibration Cell Constant Calibration Calibration by entering the cell constant with display of the conductivity 1 Press the Start softkey 2 Immerse the sensor in the solution 3 Modify the value of the cell constant until the temperature corrected conductivity value is reached Then press E 4 Finally you can Apply these values or Cancel the calibration Conductivity sensor Cell constant CLS16D k ...

Page 30: ...een replaced 5 Select TAG to enter a measuring point which is to be saved in the sensor together with the calibration data record 6 Perform the selected calibration as described on the following pages Follow the instructions on the display Calibration in Air Calibrating the slope in air 1 Place sensor in air and wait for a stable measured value 2 Press Start softkey 3 Adjust the correct value for ...

Page 31: ...t softkey Calibration will be performed 3 Finally you can Apply these values or Cancel the calibration Data Entry Calibration Calibration by entering known sensor values 1 Press Start softkey 2 Adjust the known sensor values for zero and slope 3 Finally you can Apply these values or Cancel the calibration Note To abort calibration you can press at any time Calibration ...

Page 32: ...he sensor in the medium to be measured 4 Watch the display and wait for the reading to stabilize Toggling the Measured Value Display During measurement you can toggle between display of primary secondary measured value and clock by pressing Adjusting the Temperature When you connect a sensor without temperature detector you can manually adjust the temperature for measurement or calibration 1 Press...

Page 33: ... time and threshold controlled logging The data logger records up to 10 000 entries which can be assigned to different points of measurement TAGs and notes The following data will be recorded meas point note sensor ID serial number of sensor Memosens primary value temperature time stamp device status It is always the currently selected process variable which is recorded Data logger activated Numbe...

Page 34: ...nd then save it Difference When the delta range process variable and or temperature related to the last entry is exceeded a new entry is created and the delta range is displaced upwards or downwards by the delta value The first entry is automatically created when the data logger is started Example First entry 7 Delta 1 Interval time controlled In the Interval mode the data are cyclically recorded ...

Page 35: ...ured value remains within the delta range logging is performed at the preset interval The first DIFF entry is automatically created when the data logger is started Limit value combined When one of the two limit values Min Max is exceeded the data are logged as defined by the event interval Additionally the last ten measured values before an event are recorded pre trigger As long as the measured va...

Page 36: ...key Data logger Configuring the Data Logger Prerequisite Data logger is stopped The Data logger menu shows the number of occupied entries as well as the number of free entries Configuration can also be done in the Configuration menu under Data logger 1 Press Menu softkey 2 Select Data logger and confirm by pressing E 3 Select Configure data logger and confirm by pressing E 4 Configure data logger ...

Page 37: ...0 mS cm 1 0 µS cm Delta MΩcm 0 9 999 MΩcm 1 0 MΩcm Delta salinity 0 00 45 0 g kg 1 0 g kg Delta TDS 0 00 2000 0 mg l 1 mg l Delta saturation 0 200 Air 1 Air Delta conc 0 20 0 mg l 1 mg l Delta mbar 0 1000 mbar 1 mbar 2nd difference On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval see logger type interval Difference see logger type difference Limit value Interval Basis 00 00 0...

Page 38: ...nu softkey 2 Select Data logger using the arrows and confirm by pressing E 3 Select Delete logger data using the arrows and confirm by pressing E 4 Select deletion mode Complete Data Meas point or Filter you can filter for measuring point parameter or time 5 Press Delete softkey The data are deleted according to the configuration 6 Press Back softkey to return to menu selection Starting Stopping t...

Page 39: ...r data Measured temperature Date and time Primary mea sured value Scroll using arrow keys Example 2 Curve characteristic Entry 2 of 3 Select entry using arrows Data Logger Meas point ABC Note Sensor Analog Data Logger Return Graphic Time Return Measured values Date 39 pH Oxy Cond ...

Page 40: ......

Page 41: ... icon provides information on the sensor condition maintenance request Measurement can still be performed After a calibra tion the corresponding Sensoface icon friendly neutral sad is shown togeth er with the calibration data Otherwise Sensoface is only visible in measuring mode Help text for error 21 Example of an error message Press E and Help to access the help text Hold value Menu No sensor co...

Page 42: ...ss impedance Info 05 Zero Slope Info 06 Response time too long Info 07 Operating point ISFET Info 08 Leakage current ISFET Info 09 ORP offset Info 10 Polarization Info and Help Texts When an error or status message appears on the screen proceed as follows to view the corresponding info or help text 1 Press E 2 Press the Help softkey 3 The help text will be displayed In most cases you can remedy th...

Page 43: ...ce is blocked ERR 8 Identical buffers Use a buffer solution with a different nominal value before starting the next calibration step ERR 10 Buffers interchanged Repeat calibration ERR 11 Unstable value Drift too high Leave the sensor in the liquid until the measured value is stable If this does not help replace the sensor ERR 14 Time and date invalid Set the date and time ERR 18 System error Resta...

Page 44: ...tal pH sensor Ceraliquid CPS42D Digital ORP sensor Ceragel CPS71D Digital ORP sensor Ceragel CPS72D Digital ORP sensor Memosens CPS76D Combined digital pH ORP sensor Orbipore CPS91D Digital pH sensor Orbipore CPS92D Digital ORP sensor Memosens CPS96D Combined digital pH ORP sensor Ceramax CPS341D Digital non glass pH sensor Tophit CPS441D Digital non glass pH sensor Tophit CPS471D Digital non glas...

Page 45: ...eference material of the National Institute of Standards and Technol ogy NIST in accordance with DIN 19266 are used as secondary reference buffer solutions Buffer sets CPY20 Quantity pH 2 00 5000 ml 100 ml 250 ml pH 4 01 5000 ml 100 ml 250 ml 18 ml pH 6 98 5000 ml 100 ml 250 ml 18 ml pH 9 95 5000 ml 100 ml 250 ml pH 11 87 5000 ml 100 ml 250 ml Accessories for pH Item Memosens data cable CYK20 CPY7...

Page 46: ...remains connected to the meter Conductivity Product Line Conductivity Sensors Product Name Link to Product Configurator Sensor type Condumax CLS15D Conductivity sensor Condumax CLS16D Conductivity sensor Condumax CLS21D Conductivity sensor The Product Configurator can be accessed at www endress com product name ...

Page 47: ... and DKD for comparative measurement in ultrapure water up to max 20 μS cm Please visit our website for more information on our product range www endress com Conductivity Calibration Solutions CLY11 CLY11 A 74 μS cm reference temp 25 C 77 F 500ml 16 9 fl oz CLY11 B 149 6 μS cm reference temp 25 C 77 F 500ml 16 9 fl oz CLY11 C 1 406 mS cm reference temp 25 C 77 F 500ml 16 9 fl oz CLY11 D 12 64 mS c...

Page 48: ...1C2 1 5 meters Please visit our website for more information on our product range www endress com Oxygen Product Line Oxygen Sensor Product Name Link to Product Configurator Oxymax COS22D digital oxygen sensor The Product Configurator can be accessed at www endress com product name ...

Page 49: ... display with white backlighting Keypad E 2 context sensitive softkeys Data logger 10 000 memory locations Recording Manual interval or event controlled with management of tag numbers and notes Cal data logger MemoLog Memosens only Up to 100 Memosens calibration records can be saved Viewable on the display Manufacturer sensor type serial no zero slope calibration date Temperature input 2 x 4 mm di...

Page 50: ... EMC EN 61326 1 General Requirements Emitted interference Class B residential area Immunity to interference Industry EN 61326 2 3 Particular Requirements for Transmitters RoHS conformity According to directive 2011 65 EU Power supply 4 x AA alkaline batteries or 4x rechargeable NiMH batteries 1x Li ion battery USB chargeable Nominal operating conditions Ambient temperature 10 55 C 14 130 F Transpo...

Page 51: ...ng to EN 60746 1 at nominal operating conditions 2 1 count 3 plus sensor error 4 Ranges depending on Memosens sensor Memosens pH input also ISFET M8 socket 4 pins for Memosens lab cable or M12 socket for Memosens sensors Display ranges 4 pH 2 00 16 00 mV 1999 1999 mV Temperature 50 250 C 58 482 F Memosens input ORP M8 socket 4 pins for Memosens lab cable or M12 socket for Memosens sensors Display ...

Page 52: ...0 05 NIST technical 1 68 4 00 7 00 10 01 12 46 06 NIST standard 1 679 4 006 6 865 9 180 07 HACH 4 01 7 00 10 01 12 00 08 WTW techn buffers 2 00 4 01 7 00 10 00 09 Hamilton 2 00 4 01 7 00 10 01 12 00 10 Reagecon 2 00 4 00 7 00 9 00 12 00 11 DIN 19267 1 09 4 65 6 79 9 23 12 75 Permissible calibration range Zero point 6 8 pH With ISFET Operating point asymmetry 750 750 mV Slope Approx 74 104 possibly...

Page 53: ...isplay resolution autoranging Conductivity 0 001 μS cm c 0 05 cm 1 0 01 μS cm c 0 05 0 2 cm 1 0 1 μS cm c 0 2 cm 1 Resistivity 00 00 99 99 MΩ cm Salinity 0 0 45 0 g kg 0 30 C TDS 0 1999 mg l 10 40 C Sensor standardization Cell constant Input of cell constant with simultaneous dis play of conductivity value and temperature Input of solution Input of conductivity of the calibration solution with sim...

Page 54: ...ation 000 μg l 20 00 mg l Partial pressure 0 0 1000 mbar Temperature meas range 4 20 150 C Sensor standardization Automatic calibration in air 100 RH Zero calibration Storage in quiver with moistening sponge 1 according to EN 60746 1 at nominal operating conditions 2 1 count 3 plus sensor error 4 Ranges depending on Memosens sensor ...

Page 55: ...e increasing 36 Battery replacement 12 Benchtop stand 10 Buffer set pH configuration 19 Buffer sets CPY20 45 Buffer solutions 45 C Calibration Cond auto 28 Calibration Cond cell constant 29 Calibration Cond entry of solution 29 Calibration conductivity 28 Calibration in air Oxy 30 Calibration Oxy 30 Calibration Oxy data input 31 Calibration Oxy zero point 31 Calibration pH auto 26 Calibration pH d...

Page 56: ...necting a sensor 13 Connecting cable for Memosens 13 Connections 13 Connection USB battery 12 Curve characteristic data logger 39 D Data input Oxy calibration 31 Data input pH calibration 27 Data logger configuration 36 Data logger icons 33 Data logger operating modes 34 Data logger starting 38 Data logger stopping 38 Data of the meter 49 Deleting the logger data 38 Delta range data logger 34 Devi...

Page 57: ...ductivity calibration 29 ERROR error codes 43 Error messages overview 43 EU Declarations of Conformity 7 Event interval limit value logger type 35 F Features 8 H Help texts 42 Hook 10 I Icons 14 Icons for data logger 33 In air Oxy calibration 30 Increasing the battery life 36 Information menu 15 Info texts 42 Inserting the batteries 12 Interval and difference logger type 35 Interval logger type 34...

Page 58: ...acement 30 MemoLog Memosens only 17 Memosens connecting cable 13 Memosens digital sensors 9 Memosens sensors 13 Memosens sensors conductivity 46 Memosens sensors pH 44 Messages Information menu 17 Micro USB port 13 N Note data logger 37 O Operating modes of data logger 34 Operating time of sensor sensor information 15 ORP correction sensor information 15 ORP sensors product line 44 Overview 8 Over...

Page 59: ... Programmed buffers description 9 Protective cover 10 Q Quickstart guides 7 R Rating plate 10 Rechargeable battery Li ion 12 Redox sensors product line 44 Registered trademarks 3 Replacing the batteries 12 Replacing the electrolyte Oxy 30 Replacing the membrane Oxy 30 Reservoir for KCI electrolyte 45 Return of products under warranty 3 S Safety instructions 7 Saving a membrane module replacement O...

Page 60: ...mation 15 Softkey 11 Specifications 49 Specific test report 7 Starting the data logger 38 Status messages overview 41 Stopping the data logger 38 Suspending the meter 10 Switching on the meter 14 Switching the measured value display 32 Symbols in display 14 T Table of error messages 43 Table of info texts 42 TAG point of measurement 17 TAG saving Oxy calibration 30 Technical data 49 Temperature ma...

Page 61: ...61 U USB port batttery 12 USB port micro 13 V Value added features 9 Viewing the logger data 38 W Warranty 3 Z Zero calibration Oxy 31 Index ...

Page 62: ...62 ...

Page 63: ......

Page 64: ...endress com TA 209 7MU EHE01 20150413 Software version 1 x 71265350 088995 www addresses endress com Endress Hauser Conducta GmbH Co KG Dieselstraße 24 70839 Gerlingen Germany Phone 49 7156 209 790 Fax 49 7156 28 158 ...

Reviews: