Operation
Liquiline System CA80TN
58
Hauser
3. Enter the new password again.
The password has been changed successfully.
Press the navigator for longer to return to the measuring mode.
10.3 Analyzer
Menu/Setup/Analyzer
Function
Options
Info
Mode
Read only
Manual, automatic or fieldbus
Device tag
Customized text,
32 characters
Factory setting
Analyzer_serial number
‣
Select any name for your analyzer. Use the
TAG name for example.
Measuring parameter
Read only
Range switch
Selection
• Automatic
• Manual
Factory setting
Manual
Automatic
• CA80TN-AA/CA/GR
• No options available for measuring range or
dilution
• The device uses all the available ranges
• Dilution is not displayed
• CA80TN-HR
• No options available for measuring range or
dilution
• The device uses all the internal ranges,
starting with the smallest range up to the
range that can measure the limit
concentration
• Set the limit concentration:
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Upper range
limit
• Dilution is not displayed
Manual
• CA80TN-AA/CA/GR
• Measuring range can be selected from a list
• Internal dilution is displayed in the list for
every range
• Dilution factor is displayed again separately
• CA80TN-HR
• The measuring range cannot be selected
• The internal measuring range is limited by
the limit concentration entered (
Upper
range limit
)
• Dilution can be selected from a list
Measuring range
Range switch
=
Automatic
No other options
Range switch
=
Manual
The options available depend on the device version ordered.
Internal dilution
Menu only available for device
versions
CA80TN-HR
Options depend on the
parameter
‣
Select the dilution factor from the list.
The internal measuring range changes
depending on the selected dilution. In
the measuring range displayed, only
the lower limit changes, as the upper
limit continues to be defined by the
limit concentration.