Liquiline System CA80TN
Technical data
Hauser
155
Repeatability of sensor
inputs
→ Documentation of the connected sensor
Measuring interval
Continuous (approx. 55 min), adjustable 45 min to 24 h
Digestion time
3 to 120 min
Sample requirement
Without dilution module
27 ml (0.91 fl oz)
With dilution module
• Sample: 15 ml (0.51 fl oz)
• Dilution water: approx. 20 ml (0.68 fl.oz) per measurement
Use demineralized water for expected sample concentrations of < 50 mg/l (ppm).
Reagent requirement
•
RB
1.35 ml (0.05 fl oz) per reagent and measurement
•
RK
0.9 ml (0.03 fl oz) per reagent and measurement
• Given a measuring interval of 1 hour, one set of reagents lasts for 3 months
Standard requirements
Given a calibration interval of 96 h approx. 330 ml (11.2 fl.oz) per month
Calibration interval
4 days
Cleaning interval
4 days
Maintenance interval
Once weekly, depending on the application
Maintenance effort
• Weekly: visual inspection
• Monthly: 1 hour
15.8 Environment
Ambient temperature
range
+5 to +40 °C (41 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Relative humidity
10 to 95 %, non-condensing
Degree of protection
IP55 (cabinet, analyzer stand), TYPE 3R (cabinet)