Diagnostics and troubleshooting
Liquiline System CA80PH
144
Hauser
11.6 Overview of diagnostic information
11.6.1 Device-specific, general diagnostic messages
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current
202 Selftest active
F
On
Off
Self-test active
216 Hold active
C
On
Off
• Hold active, wait
• Operating status with changed measured value
output
• Output values and status of all channels are on
hold
241 Firmware incomp.
F
On
On
Firmware error - internal
1. Update the software
2. Replace the backplane
3. Contact the service department and quote
the number displayed
242 Firmware incomp.
F
On
On
Firmware incompatible
1. Check firmware version
2. Update firmware
243 Firmware failure
F
On
On
Internal device error
1. Update the software
2. Replace displayed assembly
3. Contact the service department and quote
the number displayed
261 Electronics module
F
On
On
Electronics module defective
1. Restart the device
2. Check module
3. Replace module
262 Module connection
F
On
On
Electronics module not communicating
1. Check the module connection
2. Replace electronics module
3. Check the internal connecting cable to the
module
263 Incomp. detected
F
On
On
Incompatibility detected
The device configuration does not match the
configuration.
1. Check device settings
2. Check electronics module type
284 Firmware update
M
On
Off
Firmware update successfully completed