Commissioning
Liquiline System CA80AL
52
Hauser
4. Check that the hoses of the sample preparation system are mounted correctly in the
hose glands. It should not be possible to remove the hoses without the application of
force.
5. Visually inspect all the hose connections to ensure everything is correct.
L
WARNING
Connection errors
The safety of people and of the measuring point is under threat. The manufacturer does
not accept any responsibility for errors that result from failure to comply with the
instructions in this manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Instrument status and specifications
‣
Are the hoses free from damage on the outside?
Visual inspection of the liquid-bearing lines
‣
Check the hose connections using the hose connection diagram.
‣
Is the suction line connected to the sample collector (if present)?
‣
Are the dispensers inserted correctly?
‣
Can the dispensers move up and down freely?
‣
Are all the hose connections leak-tight?
‣
If sample preparation is provided: has the connection been made? Are the protective
hoses in the hose glands strain-relieved?
‣
If sample preparation is not connected: is the sample hose in the hose gland strain-
relieved?
‣
Have the bottles with reagents, cleaner and standard been inserted and connected?
9.3
Switching on the measuring device
1. Connect the power supply.
2. Wait for the initialization to finish.
9.4
Setting the operating language
Configuring the language
‣
Press the soft key
MENU
. Set your language in the top menu item.
The device can now be operated in your chosen language.
9.5
Configuring the measuring device
9.5.1
Basic setup analyzer
Making basic settings
1. Switch to the
Setup
/
Basic setup analyzer
menu.
Make the following settings.
2.
Device tag
: Give your device any name of your choice (max. 32 characters).
3.
Set date
: Correct the set date if necessary.
4.
Set time
: Correct the set time if necessary.