Operation
Liquiline Mobile CML18
32
Hauser
8
Operation
8.1
Calibration
To calibrate sensors, the calibration settings are first configured via the SmartBlue app. The
calibration can then be started from the device.
Configure the calibration settings in the SmartBlue app:
1. Enable Bluetooth. → 26
2. Link the device to a mobile terminal via the SmartBlue app. → 20
3. Select the device in the SmartBlue App.
4. Navigate to:
Calibration settings
Application
>>
Sensor
>>
Advanced settings
>>
Calibration settings
5. Configure the calibration settings.
e.g. Manufacturer and calibration buffer.
Perform calibration on the device:
1. Navigate to:
Guidance
Select the desired calibration.
2. Press to navigate through the calibration.
The following calibrations can be performed:
Type of calibration
Measuring parameters
Navigate to:
1-point calibration
ORP
>>
1 point calib.
2-point calibration
pH or ISFET
>>
2 point calibration
Cell constant calibration
Inductive/conductive conductivity
>>
Cell constant
Installation factor calibration
Conductive conductivity
>>
Installation factor
Air 100%rH calibration
Oxygen
>>
Air 100% rh
Air variable calibration
Oxygen
>>
Air variable
1-point calibration
Oxygen
>>
1 point calib.
8.2
Reading measured values
Measuring screens are shown on the display when a sensor is connected.
For each sensor, there are 3 measuring screens with different measured variables→ 19.
To scroll through the measuring screens:
‣
Press .
After the last measuring screen, the display returns to the first measuring screen.