iTEMP TMT71
Operating menu and parameter description
Hauser
73
14.2 Menu: Application
14.2.1 Submenu: Measured values
14.2.2 Submenu: Sensor
Unit
Navigation
Application → Sensor → Unit
Description
Use this function to select the engineering unit for all the measured values.
Selection
• °C
• °F
• K
• Ω
• mV
Factory setting
°C
Additional information
Please note: If another unit has been selected instead of the factory setting (°C), all
the set temperature values are converted to correspond to the configured
temperature unit.
Example: 150 °C is set as the upper range value. Following the selection of °F as the
engineering unit, the new (converted) upper range value = 302 °F.
Sensor type
Navigation
Application → Sensor → Sensor type
Description
Use this function to select the sensor type for the sensor input.
Please observe the terminal assignment when connecting the sensors. → 19
Selection
A list of all the possible sensor types is provided in the 'Technical data' section. → 46
Factory setting
Pt100 IEC751
Connection type
Navigation
Application → Sensor → Connection type
Prerequisite
An RTD sensor or a resistance transmitter must be specified as the sensor type.
Description
Use this function to select the connection type for the sensor.