Wiring
10
Hauser
4
Wiring
WARNING
!
Device is energized
Improper connection can cause injury or death.
•
The electrical connection must only be carried out by a certified electrician.
•
Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
•
Prior to beginning
any wiring work, make sure voltage is not applied to any of the cables.
4.1
Electrical connection to the transmitter
NOTICE
Usage of impermissible cable diameter
Losing the protection class IP65
•
Use a cable with a diameter of 6 ... 12 mm (0.24 ... 0.47") for Pg 13.5 cable gland.
•
Use a cable with a diameter of 13 ... 18 mm (0.51 ... 0.71") for Pg 21 cable gland (designated for
mains).
Connect the cleaning unit to the CM44x transmitter by applying the voltage L1 to terminals 41 and 42.
a0016124
Fig. 4: Connection