Commissioning
Proline Promag P 500 HART
98
Hauser
10.4.3 Displaying the I/O configuration
The
I/O configuration
submenu guides the user systematically through all the parameters
in which the configuration of the I/O modules is displayed.
Navigation
"Setup" menu → I/O configuration
‣
I/O configuration
I/O module 1 to n terminal numbers
I/O module 1 to n information
I/O module 1 to n type
Apply I/O configuration
Alteration code
Parameter overview with brief description
Parameter
Description
User interface / Selection /
User entry
Factory setting
I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O
module.
• Not used
• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4)
–
I/O module 1 to n information
Shows information of the plugged I/O
module.
• Not plugged
• Invalid
• Not configurable
• Configurable
• HART
–
I/O module 1 to n type
Shows the I/O module type.
• Off
• Current output
• Current input
• Status input
• Pulse/frequency/switch
output
Off
Apply I/O configuration
Apply parameterization of the freely
configurable I/O module.
• No
• Yes
No
Alteration code
Enter the code in order to change the I/O
configuration.
Positive integer
0
*
Visibility depends on order options or device settings
10.4.4 Configuring the status input
The
Status input
submenu guides the user systematically through all the parameters that
have to be set for configuring the status input.