Installation
Cerabar PMP71B HART
26
Hauser
1
A0043807
1
Locking screw
NOTICE
The housing cannot be unscrewed fully.
‣
Loosen the external locking screw by a maximum of 1.5 turns. If the screw is
unscrewed too much or completely (beyond the screw anchor point), small parts
(counter disk) can become loose and fall out.
‣
Tighten the securing screw (hex socket 4 mm (0.16 in)) with maximum
3.5 Nm (2.58 lbf ft) ± 0.3 Nm (0.22 lbf ft).
5.3
Post-installation check
Is the measuring device undamaged (visual inspection)?
Are the measuring point identification and labeling correct (visual inspection)?
Is the measuring device protected against precipitation and sunlight?
Are the securing screws and cover lock tightened securely?
Does the measuring device meet the measuring point specifications?
For example:
• process temperature
• process pressure
• ambient temperature
• measuring range